Calibration
Carrying out the zero calibration:
1. Start zero calibration in PolySoft.
®
 WirelessHART
interface transmits the device variable status "Bad" and
the special status "Calibration".
2. Follow the instructions of the wizard In PolySoft.
Flow with test gas supply: 0.5 L/min ± 10 %
3. After completion of the zero calibration, switch off the gas flow and remove
the calibration adapter.
11.5
Carrying out the span calibration
Prerequisites:
Zero calibration has been carried out.
The test gas supply has been prepared.
A connection between the gas detector and PolySoft has been established.
Carrying out the span calibration:
1. Start and confirm span calibration with PolySoft.
®
 WirelessHART
interface transmits the device variable status "Bad" and
the special status "Calibration".
2. Follow the instructions of the wizard In PolySoft.
Flow with test gas supply: 0.5 L/min ± 10 %
3. After completion of the span calibration, switch off the gas flow and remove
the calibration adapter.
48
Instructions for use
|
Dräger Polytron
®
6100 EC WL, Dräger Polytron
®
®
Repeater WirelessHART