Troubleshooting
Byte
Bit
Format
8
6 and 7
Bits
12.2
Command 48: Reading additional device status
Byte
Bit
Description
0
0
VOLTAGE_OUT_OF_RANGE
2
BATTERY_WARNING_LOW
6
SENSOR_WARMUP
1
1
SENSOR_VITALITY_FAIR
2
SENSOR_VITALITY_POOR
4
CALIBRATION_INTERVAL_EXCEEDED
7
DEVICE_TEMPERATURE_WARNING
6
0
MAINTENANCE_REQUIRED
1
DEVICE_VARIABLE_ALERT
2
CRITICAL_POWER_FAILURE
3
FAILURE
4
OUT_OF_SPECIFICATION
5
FUNCTION_CHECK
50
Extended device status
Description
Device variable status
Meaning of "Quality" (Bit 6 and 7):
3: Good – device variable value
with normal accuracy.
2: Poor – device variable value with
low accuracy.
1: Manual/Fixed – device variable
value fixed.
0: Bad – There may be a device
variable value, but the device vari-
able value may not be used.
Instructions for use
|
Dräger Polytron
Cause
Cause
Battery voltage below warning limit.
Sensor in warm-up phase 2. No calibration possi-
ble.
The residual sensitivity is between 25 and 50 %.
The residual sensitivity is between 0 and 25 %.
Calibration interval expired.
NTC sensor greater or smaller than permitted
temperature.
All alarms and warnings.
Battery charge critical.
®
6100 EC WL, Dräger Polytron
Remedy
Remedy
Contact DrägerService.
Change battery pack.
Wait for warm-up phase 2 to finish.
Consider a sensor change.
Consider a sensor change.
Calibrate the sensor.
In case of high temperatures due to solar radiation,
use weather protection set (accessory).
Check warning and error status.
Change battery pack immediately.
®
Repeater WirelessHART
®