Troubleshooting
Byte
Bit
Description
16
0
OUT_OF_SERVICE
1
INSTALLATION_CALIBRATION
2
FAULTS_SENSOR
3
FAULTS_ELECTRONICS
4
M: Maintenance required
5
O: Out of specification
6
C: Function Check
7
F: Failure
17
0
NO_SENSOR_ERROR
1
ERROR_CAPTURE_CURRENT
2
ERROR_SENSOR_LOCK
3
ERROR_SDS_CYCLIC_CHECK
4
PROGRAM_FLOW_ERROR
52
Instructions for use
|
Dräger Polytron
Cause
Start sequence active.
Start sequence failed.
Calibration or bump test active.
Sensor not detected or defective.
For the activated sensor lock feature, a sensor
with a different part number was inserted.
Serious device fault, various causes.
No sensor used.
Error in EC sensor.
Sensor lock feature enabled.
Error in EC sensor.
Error in electronics.
®
6100 EC WL, Dräger Polytron
Remedy
Wait for start sequence.
A
Restart the device.
B
Have the gas detector checked by DrägerService.
Finish calibration or bump test.
A
Check the sensor connection.
B
Remove and reinstall the sensor.
C
Change the sensor.
D
Insert sensor with the correct part number.
A
Restart the device.
B
Reset device to factory settings.
C
Update firmware.
D
Have the gas detector checked by DrägerService.
Use sensor.
Change the sensor.
A
Insert sensor with the correct part number.
B
Deactivate sensor lock feature.
Change the sensor.
A
Restart the device.
B
Reset the sensor to the factory settings.
®