3
Description
3.1
Product overview
Illustration on fold-out page:
A 1
Housing cover
A 2
Battery box
A 3
Protective cap (if battery box removed)
A 4
Connector for antenna with cover cap (delivery state)
A 5
Sealing ring
A 6
Screw plug
A 7
Housing
A 8
Sensing head
A 9
Bayonet ring
A 10
Dummy disk (delivery state)
A 11
Trim plate
A 12
®
Marking for Bluetooth
activation /
magnetic pen use
A 13
LED display
A 14
Set screw (Torx T20)
A 15
Catch-rope
A 16
Fastening screw for catch-rope
A 17
Fastening screw for potential equalization
Instructions for use
|
Dräger Polytron
®
6100 EC WL, Dräger Polytron
3.2
Feature description
®
3.2.1
Dräger Polytron
6100 EC WL
®
The Dräger Polytron
6100 EC WL is a fixed gas detector for monitoring toxic
gases and oxygen. The gas detector includes a repeater feature for forwarding
®
data from other WirelessHART
network users.
Operation is carried out with a battery pack or a wired power supply.
The gas detector is integrated into a WirelessHART
manager, access point, and controller. The controller is the Dräger
®
REGARD
7000 or a PLC for example. Ranges between access point and gas
detector are up to 500 m in the open.
Operation and configuration
The general device status is signaled via the LED display.
Operation and configuration are carried out with the Dräger PolySoft
configuration software. The connection with PolySoft is established via the
®
Bluetooth
interface installed.
Contact the Dräger sales team for obtaining the Dräger PolySoft
configuration software. The software can be downloaded from
www.draeger.com/software.
Operation
®
A WirelessHART
process interface communicates the gas reading and device
status via burst mode to the controller.
Alarms
The alarms configured in the gas detector are not transmitted via the
®
WirelessHART
process interface. Alarms are only output via the red LED.
They are purely informative and are used to locate a gas detector raising the
alarm and are not intended for the derivation of a safety task or action. The
alarms are non-latching. When alarm conditions are resolved, the alarm is no
longer displayed.
If alarm logic is used for countermeasures, implement alarm logic in the
controller.
®
Repeater WirelessHART
|
Description
®
network with system
®
en
37