Télécharger Imprimer la page

Microlife BP A6 BT Mode D'emploi page 7

Masquer les pouces Voir aussi pour BP A6 BT:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 24
 There is a break of 15 seconds between the measurements.
A count down indicates the remaining time.
 The individual results are not displayed. Your blood pressure
will only be displayed after all 3 measurements are taken.
 Do not remove the cuff between measurements.
 If one of the individual measurements was questionable, a
fourth one is automatically taken.
4. Taking a blood pressure measurement
Checklist for taking a reliable measurement
1. Avoid activity, eating or smoking immediately before the
measurement.
2. Sit down on a back-supported chair and relax for 5 minutes.
Keep your feet flat on the floor and do not cross your legs.
3. Always measure on the same arm (normally left). It is recom-
mended that doctors perform double arm measurements on a
patients first visit in order to determine which arm to measure in
the future. The arm with the higher blood pressure should be
measured.
4. Remove close-fitting garments from the upper arm. To avoid
constriction, shirt sleeves should not be rolled up - they do not
interfere with the cuff if they are laid flat.
5. Always ensure that the correct cuff size is used
(marking on the cuff).
 Fit the cuff closely, but not too tight.
 Make sure that the cuff is positioned 1-2 cm above the
elbow.
 The artery mark on the cuff (ca.3 cm long bar) must lie over
the artery which runs down the inner side of the arm.
 Support your arm so it is relaxed.
 Ensure that the cuff is at the same height as your heart.
6. Slide the lock switch AO down to the «unlock» position. Press
the START/STOP button 1 to start measuring.
7. The cuff will now pump up automatically. Relax, do not move
and do not tense your arm muscles until the measurement
result is displayed. Breathe normally and do not talk.
8. When the correct pressure is reached, the pumping stops and
the pressure falls gradually. If the required pressure was not
reached, the device will automatically pump some more air into
the cuff.
9. During the measurement, the pulse indicator BQ flashes in the
display.
Microlife BP A6 BT
10.The result, comprising the systolic AP and the diastolic AQ
blood pressure and the pulse rate AR are displayed. Note also
the explanations on further display symbols in this booklet.
11.When the device has finished measuring, remove the cuff.
12.Switch off the device. (The monitor does switch off automati-
cally after approx. 1 min.).
How not to store a reading
As soon as the reading is displayed press and hold the START/
STOP button 1 until «M» BO is flashing. Confirm to delete the
reading by pressing the M-button 3.
You can stop the measurement at any time by pressing the
ON/OFF button or open the cuff (e.g. if you feel uneasy or
an unpleasant pressure sensation).
If the systolic blood pressure is known to be very high,
it can be an advantage to set the pressure individually.
Press the START/STOP button after the monitor has been
pumped up to a level of approx. 30 mmHg (shown on the
display). Keep the button pressed until the pressure is
about 40 mmHg above the expected systolic value – then
release the button.
How do I evaluate my blood pressure
The LED traffic light indicator on the left-hand side of the display
7 indicates within which range the measured blood pressure lies.
The value is either within the optimum (green), elevated (yellow) or
high (red) range. The classification corresponds to the following
ranges defined by international guidelines (ESH, ESC, JSH). Data
in mmHg.
Range
Systolic Diastolic Recommendation
≥135
1. blood pressure
too high
2. blood pressure
130 - 134 80 - 84
elevated
3. blood pressure
<130
normal
The higher value is the one that determines the evaluation.
Example: a blood pressure value of 140/80 mmHg or a value of
130/90 mmHg indicates «blood pressure too high».
≥85
Seek medical
advice
Self-check
<80
Self-check
5
EN

Publicité

loading