Caution: Attention should be paid to ensure that the plug and the cable are kept
completely dry.
Note: Do not get the connections confused.
Calibration of probe
OK
Calibration of probe
pH7
OK
Calibration of probe
(20)
Caution: Do not rub the probe to dry it since this will cause electro-static charge and
may result in incorrectly measured values.
Calibration of probe
pH4
OK
Calibration of probe
(20)
Calibration of probe
successful
mV
Error value!
In the pH calibration menu item, start the mode by
pressing the OK button.
Dip the pH probe into the pH 7 buffer solution and
start by pressing the OK button.
Press
OK again to trigger the countdown
After 20 seconds, the control unit shows the value of
the used buffer solution and the slope in mV. If the
calibration
is
Then take the probe from the first buffer solution and,
if possible rinse with water and dry with a dry, lint-free
paper towel.
Now dip the probe into the second pH 4 buffer solution.
Press
OK again to trigger the countdown.
After another 20 seconds, the control unit shows the
values
of
the
If the calibration was valid, continue using OK
The display will appear with the specification of the lope
Go back using Mode or accept using OK
The control unit has thus been successfully calibrated
to the pH probe and the data saved.
If the calibration process was not successful, Error is
displayed.
Cancel
repeat using OK
45