Bromine reagent
Free bromine (HOBr, e.g. by
mixing sodium bromide +
chlorine bleach solution)
DBDMH: 1.3 dibromine-
5.5-dimethyl-hydantoin,
e.g. with trade designation
®
Albrom 100
BCDMH: 1-bromine-3-chlorine- Suitable for the entire measuring
5.5-dimethyl-hydantoin,
e.g. with trade designation
Brom-Sticks
®
N-bromine-amidosulfonate,
e.g. with the trade designation
®
Stabrex
Bromine concentration can be optimally controlled when it is used together
with the DULCOMETER
pumps. Typical applications include treatment of fresh water (also sea water)
for swimming pools, spa pools (whirlpools) and process water, e.g. power
station cooling towers.
5
Sensor design
The BRE bromine sensor consists of three main parts, the top part, the
electrode shaft and the membrane cap (see figure 1). The membrane cap
filled with electrolyte forms the measurement compartment into which the
electrodes are immersed.
A micro-porous membrane separates the measuring compartment from the
sample medium.
The amplifier electronics are embedded in plastic in the upper part of the
electrode shaft.
The sensor has a passive 4-20 mA two-wire interface. The sensor's power
supply is located in an external controller, e.g. DULCOMETER
Measured
variables
Total bromine: Free bromine (HOBr, OBr
(BCDMH, e.g. Bromsticks
bound bromine (Stabrex
Function of
sensor
The BRE bromine sensor is a membrane covered amperometric dual -
electrode sensor. A platinum cathode is used as working electrode and a
silver halogen coated anode is used as the counter- and reference electrode.
The bromine compounds contained in the sample water diffuse through the
®
ProMinent
Sensor type BRE 1-mA-10 ppm
(BRE 1-mA-2 ppm)/calibration
method
Max. concentration:
5 ppm (1 ppm)/calibratable with
DPD 1 or with DPD 4
Max. bromine concentration:
5 ppm (1 ppm)/calibratable with
DPD 1 or with DPD 4
range, calibratable only with
DPD 4
Suitable for the entire measuring
range, calibratable only with
DPD 4
®
D1C Bromine controller and ProMinent metering
Design and function
®
®
).
–
), organically bound bromine
or DBDMH, e.g. Albrom 100
Design and function
Sensor type BRE 2-mA-10 ppm/
calibration method
Suitable for the entire measuring
range/calibratable with DPD 1 or
with DPD 4
Suitable for the entire measuring
range/calibratable with DPD 1 or
with DPD 4
Min. bromine concentration:
0.4 ppm/calibratable only with
DPD 4
Min. bromine concentration:
0.4 ppm/calibratable only with
DPD 4
®
D1C Bromine.
®
) and inorganically