1.0
INTRODUCTION
➍
➋
➊
➏
➌
Figure 1
TRACHEOPORT
COMPACT
®
➊
Vacuum gauge
➋
Vacuum control
➌
On / Off button
➍
Battery capacity status indicator
➎
Operating status indicator
➏
Bacterial filter
➐
Connection for power supply unit
➑
Secretion container
➒
Connecting tube
➓
Secretion outlet with stopper
Connection for suction tube with stopperl
11
Air filter
12
1.2
FUNCTION
TRACHEOPORT
COMPACT is a particularly handy
•
®
small-size suctioning device.
The device is powered by a maintenance-free electric swing-
•
piston pump. When in operation, the pump generates a
vacuum in the tubing system and in the secretion container,
with the aid of which secretions or rinsing fluids are suctioned
off (via a sterile suction catheter).
The fluid is collected in the secretion container.
•
The final vacuum, and thus also the suction power, can be
•
adjusted with the aid of the fine control and the vacuum
gauge.
A bacterial filter prevents intrusion of bacteria into the inside
•
of the unit.
In addition, a mechanical overflow protector (float) integrated
•
on the lid of the secretion container prevents accidental
aspiration of secretions into the pump head.
TRACHEOPORT
N
• for continuous operation in connection with drainage in the low-vacuum range (e.g. thoracic drainage)
or long-lasting use in connection with endoscopy;
• outside of the medical field (for aspiration of flammable, caustic or explosive liquids/gases).
➒
➎
11
10
➐
➑
12
COMPACT must not be used:
®
1.3
PURPOSE
Name:
TRACHEOPORT
COMPACT
®
Main function:
Temporary and spontaneous aspiration of secretions, blood and
body fluids that typically arise during suctioning of the airways.
Medical indication / application:
Suctioning of the upper airways.
Specification of the main functions:
Drainage and temporary collection of body secretions. A
vacuum is generated by means of an electric suctioning pump.
The integrated secretion container enables temporary collection
of the body secretions that have been drained off.
Application organ:
Upper airways (oral cavity, nasopharyngeal space) and
bronchial system.
Duration of application:
Temporary application in patients undergoing suctioning of the
airways (operating time of the battery up to 60 minutes).
Application environment:
The application environment for airway suctioning is the clinical,
medical practice, nursing and homecare field. Application must
only be carried out by medically trained and briefed persons.
Contraindication:
Not suitable for:
continuous operation in connection with drainage in the
•
low-vacuum range (e.g. thoracic drainage or wound
drainage),
long-lasting use in connection with endoscopy,
•
use outside of the medical field,
•
aspiration of flammable, caustic or explosive substances,
•
suctioning in areas exposed to explosion hazards.
•
TRACHEOPORT
COMPACT is an active medical device.
®
Reprocessing:
The product is intended for repeated use. The accessories are
partly reusable. For information on reprocessing, cleaning and
disinfection, refer to instructions for use (see Section 6.0).
27