37.2. RECORDING TIME
[SENSe]:SWEep:TIME[?] (Command/Query)
37.3. MEASUREMENT
MEASure:DMM?
[SENSe]:FUNCtion
37.4. ERROR
SYSTem:ERRor[:NEXT]? (Query)
37.4.1. * COMMAND ERROR: (-199 TO -100)
They indicate that a syntax error has been detected by the syntax analyzer and causes event register bit 5, called CME, CoMmand
Error to be set to 1.
-101
:
Invalid character
-103
:
Invalid separator
-104
:
Data type error
-108
:
Parameter not allowed
-109
:
Missing parameter
-111
:
Header separator error
-112
:
Program mnemonic too long
-113
:
Undefined header
-114
:
Header suffix out of range
-121
:
Invalid character in number
-128
:
Numeric data not allowed
-131
:
Invalid suffix
-138
:
Suffix not allowed
-141
:
Invalid character data
-148
:
Character data not allowed
-151
:
Invalid string data
-154
:
String data too long
-171
:
Invalid expression
The SWE:TIME <time|MAX|MIN|UP|DOWN> command sets the recording time.
<time> is a value in NRf format and may be followed or not by a multiple of the unit.
By default, it is expressed in second.
To the question SWE:TIME? the instrument returns the recording time.
Response format: <time><NL>
value in the <NR3>
(Query)
To the question MEAS:DMM? <INT1|2|5> the instrument returns the value of the main measurement
for the selected channel.
INT5 is associated to power measurement.
(Command/Query)
FUNC <VOLTage|RESistance|CONTinuity|CAPAcitor|DIODe|RPM|POWer|POW3PN|POW3P>
selects the measurement function on channel 1.
To the question FUNC?, the instrument returns the measure function to channel 1.
To the question SYST:ERR?, the instrument returns the number of error positioned at the top of the
queue. The queue has a stack of 20 numbers and is managed as follows : first in, first out.
As the SYST:ERR? question arrive, the instrument returns the number of errors in order of arrival, until
the queue is empty. Every more SYST:ERR? question involves a negative answer: character "0" (ASCII
48code). If the queue is full, the case at the top of the queue takes the value -350 (saturated queue).
The queue is empty:
- when the instrument is getting started.
- at the receipt of a *CLS.
- at the reading of the last error.
Response format: <error><NL>
with error = negative or 0, no error.