36.2.2. TRIGGER MODE - AUTOMATIC MODE
TRIGger[:SEQuence
{[1]|2}]
:ATRIGger[:STATe]
36.2.3. SINGLESHOT MODE
INITiate[:IMMediate]:
NAME
36.3. HORIZONTAL
36.3.1. MIN/MAX ACQUISITION
[SENSe]:AVERage:
TYPE
36.3.2. AVERAGE
[SENSe]:AVERage:
COUNt
36.3.3. TIME BASE
DISPlay[:WINDow]
:TRACe:X[:SCALe]
:PDIVision
[SENSE]SWEep:OFFSet
:TIME
(Command/Query)
The TRIG:ATRIG <1|0|ON|OFF> command validates or devalidates the automatic trigger.
ON|1 activates the auto trigger mode
OFF|0 activates the trigger mode
To the question TRIG:ATRIG ?, the instrument returns the activation of the auto trigger.
(Command)
The INIT:NAME <{EDGE|PULse }> command launches a singleshot acquisition in the indicated trigger
mode.
(Command/Query)
The AVER:TYPE <NORMal|ENVelope> command validates or devalidates the mode of min/max
acquisition.
NORMal devalidates the mode of min/max acquisition.
ENVelope validates the mode of min/max acquisition.
To the question AVER:TYPE?, the instrument returns the activation status of the mode of min/max
acquisition.
(Command/Query)
The AVER:COUN <acquisition number|MAX|MIN|UP|DOWN> command determines the number of
acquisition bursts necessary to obtain a displayed trace by averaging.
<acquisition number> is a value in format NR1, from values 0, 2, 4, 16 to 64.
To the question AVER:COUN?, the instrument returns the number of acquisition bursts necessary to
obtain a displayed trace by averaging.
(Command/Query)
The DISP:TRAC:X:PDIV <scale|MAX|MIN|UP|DOWN > command sets the value of the time base.
<scale> is a value in format NRf, it may be followed or not by a multiple and by the unit.
By default, the value is expressed in second.
Example: to get a time base of 1 µs, following values can be entered: 1E-3ms or 1E-6 or 0.000001s or
0.000001 or else 1us
To the question DISP:TRAC:X:PDIV?, the instrument returns the value of the time base.
Response format:
value in format <NR3> expressed in second.
(Command/Query)
The SWE:OFFS:TIME <time|MAX|MIN|UP|DOWN> command sets the horizontal offset of the trace
(run-after-delay or postrig).
<time> is a signed value in format <NRf> ; it may be followed or not by a multiple and by the unit.
By default, it is expressed in second.
To the question SWE:OFFS:TIME?, the instrument returns the current run-after-delay.
Response format: <measured value><NL>
value in format <NR3> expressed in second.
<measured value><NL>