Flow Range:
Low Flow Alarm:
Average Power
Consumption:
Pressure Relief Mechanism
Operational at:
Change in maximum
recommended flow when
back pressure of 7kPa is
applied:
Filters:
Safety System:
Width:
Height:
Depth:
Weight:
Shipping Weight:
Operating Ambient
Temperature:
Exhaust Cooling Air
Temperature:
Oxygen Output
Temperature:
Cabinet:
Part No 1163146
SECTION 5—TYPICAL PRODUCT PARAMETERS
0.5 to 5 L/min (maximum). For flowrates less
than 1 L/min, we recommend the use of the
Invacare Pediatric Flowmeter Accessory
(IRCPF16AW).
0 L/min to 0.5 L/min
Rapid Audible Alarm Beeping (No
Accessories Connected)
300 W
280 W @ 3 L/min
241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)
0.7 L/min
Cabinet, Outlet HEPA and Compressor Inlet
Current overload or line surge shutdown. High
temperature compressor shutdown. High
Pressure Alarm w/compressor shutdown. Low
Pressure Alarm w/compressor shutdown.
Battery Free Power Loss Alarm. SensO
System (SensO
38.1 cm ± 1 cm (15 in ± 3/8 in)
58.4 cm ± 1 cm (23 in ± 3/8 in)
30.5 cm ± 1 cm (12 in ± 3/8 in)
20.5 kg ± 1 kg (45 lbs ± 2 lbs)
22.7 kg ± 1 kg (50 lbs ± 2 lbs)
10°C - 35°C (50°F - 95°F) at 20-60% relative
humidity
Less than Ambient +19°C (+ 35°F)
Less than Ambient +3°C(+ 6°F)
Impact Resistant flame-retardant plastic cabinet
that conforms to UL 94-V0.
>
Model) Low Flow Alarm.
2
15
Oxygen
2
Perfecto
™Series
2