Proline Multichannel Adjust able Volume Pipet t es
Cat.No.
Volume
Volume
range/µl
µl
720210
8- ch
10
0.5-10
5
1
720220
8- ch
50
5-50
25
5
720240
8- ch
300
50-300
150
50
720310
12-ch
10
0.5-10
5
1
720320
12-ch
50
5-50
25
5
720340
12-ch
300
50-300
150
50
Proline Fixed Volume Pipet tes
Cat.No.
Volume
Volume
range/µl
µl
722001
5
5
722004
10
10
722010
20
20
722015
25
25
722020
50
50
722025
100
100
722030
200
200
722035
250
250
722040
500
500
722045
1000
1000
722050
2000
2000
722055
5000
5000
Liquid:
Distilled water (grade 3, ISO 3696)
Reference temperature: 22°C. constant to ± 0.5°C
T ested:
According to ISO 8655 using original
Sartorius Optifit Tips
Inaccuracy
Imprecision
µl
µl
1.50
1.50
2.50
2.50
4.00
4.00
1.00
0.50
1.50
1.00
3.00
2.00
0.70
0.25
1.00
0.50
1.50
0.80
1.50
1.50
2.50
2.50
4.00
4.00
1.00
0.50
1.50
1.00
3.00
2.00
0.70
0.25
1.00
0.50
1.50
0.80
Inaccuracy
Imprecision
µl
µl
1.30
1.20
0.80
0.80
0.60
0.50
0.50
0.30
0.50
0.30
0.50
0.30
0.40
0.20
0.40
0.20
0.30
0.20
0.30
0.20
0.30
0.15
0.30
0.15