CAUTION
Be sure to clean and lubricate the handpiece prior to autoclaving. If blood remains on the internal surface it can become
clotted and cause product failure.
Do not autoclave the product with other instruments even when it is in a pouch. This is to prevent possible discoloration and
damage to the product from chemical residue on other instruments.
Do not heat or cool the product too quickly. Rapid change in temperature could cause damage to the product.
To avoid product failure, do not use a sterilizer whose temperature exceeds 138°C during use including the dry cycle.
The chamber temperature of some type of sterilizers may exceed 138°C. Contact the sterilizer manufacturer for detailed
information.
Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight. The air should be free
from dust, salt and sulphur.
Do not touch the product immediately after autoclaving as it will be very hot and must remain in a sterile condition.
Autoclave sterilization is recommended for the product. The validity of other sterilization methods is not confirmed.
Only Handpiece, Internal Spray Nozzle and Nozzle Holder can be autoclaved. Do not autoclave other components (Y-Connector,
Irrigation Tube, etc.).