8-1 Preparation
1) Wear a eye protection, a mask and gloves to avoid infection.
2) Remove the bur.
3) (Only for X-SG20L / SG20) Immerse about 20mm of the handpiece head in water,
and rotate the handpiece for about 15 seconds while switching the Forward/Reverse
rotation direction (Fig. 11)
4) Disconnect the handpiece from the motor.
8-2 Cleaning
Manual Cleaning (External Surface)
1) Remove dirt and debris on the handpiece surface using a brush under running water (at
38°C or less). Do not use a wire brush.
2) Wipe off the moisture on the surface using absorbent cloth.
CAUTION
If water comes inside the handpiece, blow air into the rear of the handpiece using an air syringe and remove the remaining
water.
Fig. 11