TROUBLESHOOTING
The device switches
off during operation.
The device
does not switch off
automatically
even though the
medicament
holder is empty.
The medicament
holder leaks.
The inhalation with
the mask takes longer
If the interference cannot be remedied, please contact the manufacturer before trying to
repair the device on your own.
The medicament holder is not
attached correctly.
The electrodes at the medicament
holder are dirty.
The power adapter has no proper
contact to the device.
The medicament holder is empty.
There has been no contact for
more than 10 seconds
between the inhalation solution
and the mesh membrane.
The device was shaken vigorously
during operation.
The medicament holder is
defective.
The inhalation solution used
strongly foamed up.
The electrodes at the medicament
holder are dirty.
The electrodes at the main device
and/or the medicament holder
are discoloured.
The medicament holder is
defective.
The medicament holder is
defective.
The sealing of the medicament
holder is defective.
This has technical reasons.
.
Through the mask openings,
you inhale less inhalant per breath
than via the mouthpiece. The
inhalant is mixed with ambient air
through the holes.
Attach the medicament holder
correctly and restart the device.
Clean the electrodes and restart
the device.
Connect the power adapter
correctly.
Fill the medicament holder.
Slightly tilt the inhaler in order to
establish a contact between
the inhalation solution and the mesh
membrane.
Calmly and firmly hold the device
during operation in your hand.
Replace the medicament holder by
a new one.
Clean the medicament holder and
restart the device.
Clean the electrodes and restart
the device.
Clean the electrodes and restart
the device.
Replace the medicament holder
by a new one.
Replace the medicament holder
by a new one.
29
DE
EN
FR
IT
ES
NL
RU
PL
TR
AR