TROUBLESHOOTING
TROUBLESHOOTING
Extremely low
nebulisation
After switching on
the operating mode
LED flashes for one
second and then
turns off immediately.
DE
EN
FR
The operating mode
LED does not light up
and the device
does not start.
IT
ES
The operating mode
LED lights up and
NL
the device does
not start.
RU
PL
TR
AR
The electrodes at the medicament
holder are dirty
The electrodes at the main device
and/or the medicament holder are
discoloured.
The medicament holder
is not attached correctly.
The medicament holder is empty.
There has been no contact for
more than 10 seconds between the
inhalation solution and the mesh
membrane.
The electrodes at the main device
and/or the medicament holder are
discoloured.
The batteries are inserted with
wrong polarity.
The batteries are empty.
The power adapter is wrongly
connected.
The battery status LED lights up
and shows that the batteries are
empty.
The mesh membrane is ripped.
The electrodes at the medicament
holder are dirty.
The electrodes at the main device
and/or the medicament holder
are discoloured.
The mesh membrane is dirty.
Clean the electrodes and restart
the device.
Clean the electrodes and restart
the device.
Attach the medicament holder
correctly and restart the device.
Fill the medicament holder.
Slightly tilt the inhaler in order to
establish a contact between the
inhalation solution and the mesh
membrane.
Clean the electrodes and restart
the device.
Insert the batteries with the right
polarity.
Replace the old batteries by new
ones.
Connect the power adapter
correctly.
Replace the old batteries by new
ones.
Replace the medicament holder
by a new one.
Clean the electrodes and restart
the device.
Clean the electrodes and restart
the device.
Clean the mesh membrane. If it still
does not work after cleaning,
replace the medicament holder
by a new one.