Switching Off The Device; Troubleshooting; Corrective Action - ProMed INH-2.1 Instructions D'utilisation

Inhalateur
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 36
OPERATING THE DEVICE
Caution!
Ethereal medicinal plant oils, cough syrups, gargle solutions, drops for embrocation or steam baths
are generally unsuitable for the inhalation with inhalers. These additives are often viscous and may
adversely affect the proper function of the device and thus the effectiveness of the application. In
the event of a hypersensitivity of the bronchial system, medicaments with ethereal oils may possibly
trigger an acute bronchospasm (a sudden spasmodic construction of the bronchial tubes with
difficulty in breathing). Please contact your doctor or pharmacist!
G. Switching off the device
To switch off the device, briefly press the Start/Stop button; the inhaler switches off and the operating
mode LED goes out. When the content of the medicament holder is used up, a short high-frequency
signal sounds and the device automatically switches off.
Note:
If you use a power adapter (not supplied with the product, can only be purchased as accessory), you
have to remove it from the device after having switched it off.
H. Using a power adapter
If you use a power adapter, insert the DC connector
of the adapter into the power supply connection socket (Q) first and then the power adapter into the
socket.
Note:
1.
Only use power adapters which are approved for the Promed INH-2.1 (Fuhua, UE05WCP-
030100SPC).
2.
To avoid possible damages, you have to disconnect the power adapter from the device after
use.
3.
The power supply unit can easily heat up. This does not indicate a malfunction. The device
can be used without problems.
4.

Troubleshooting

Error/
error display
Extremely low
nebulisation
Cause
The medicament holder
is not completely
connected.
There has been no contact for
more than 10 seconds
between the inhalation solution
and the mesh membrane.
The mesh membrane is blocked.
27

Corrective action

Connect the
medicament holder
correctly and restart
the device.
Slightly tilt the inhaler in order to
establish a contact between the
inhalation solution and the mesh
membrane.
Clean the medicament holder.
If it still doesn't work after cleaning,
replace the medicament holder
by a new one.
DE
EN
FR
IT
ES
NL
RU
PL
TR
AR

Publicité

Table des Matières
loading

Table des Matières