OPERATING THE DEVICE
Caution!
To avoid malfunctions and damages, ensure that the medicament holder is correctly clicked into
place.
Caution!
Please keep the contact pins (L) of the main device and the medicament holder clean. Otherwise, the
device cannot work properly.
Caution!
Keep the small part away from children as they may inhale or swallow it.
D. Attaching the mouthpiece or an inhalation mask
First of all, remove the protective cap (C) from the medicament holder and attach either the
mouthpiece (D) or an inhalation mask (E).
Note!
Under normal conditions, the lifetime of the mouthpiece is approx. 2 years (when using the device
three times per day or 30 minutes per day).
E. Switching on the device
To switch on the inhaler, briefly press the Start/Stop button (F) once; the
operating mode LED (G) lights up green. When there is no liquid in the
medicament holder, the operating mode LED lights up green briefly and the
device switches off automatically. After switching on, the device starts the
nebulisation for approx. 1 second, pauses for approx. 1/2 second and then
starts the continuous operation.
Note:
1.
You can carry out a functional test of the device by filling the medicament holder with a 0.9%
sodium chloride (NaCl) solution and starting the device.
2.
The device cannot be operated with purified water. If there is only purified water in the
medicament holder, the device cannot be switched on.
Caution!
1.
If you hold down the Start/Stop button when switching on the device, the device automatically
switches into the cleaning mode and the operating mode LED lights up blue. The cleaning mode
must not be used for inhaling.