3. Installation
Connect the mains plug to a suitable socket outlet. Make sure the mains
plug is easily accessible so it can be disconnected quickly in an
emergency.
When the bed is connected to the electricity supply, indicators will light on
the control box (1) and Attendant Control Panel.
The power supply cord (2) is fitted with a plastic hook (3). When not in
use or before moving the bed, clip the hook onto the head board, coil up
the cable and place it over the hook as shown.
To isolate the bed from the electricity supply, disconnect the mains plug
from the socket outlet.
2
Cable storage and potential equalisation terminal
A potential equalisation terminal (4) is located at the head end of the
bed.
When other electrical equipment is within reach of the patient or carer,
potential differences between the equipment can be minimised by
connecting together their potential equalisation terminals.
3
1
13
4