5.5 Setting up the scale

Aligning the scale
2 .
Spirit level
Air bubble
74 •
1. Place the scale on firm, level surface.
2. Undo the knurled wheels.
3. Level the device by turning the foot screws.
4 .
3 .
The air bubble in the spirit level must be located in
the exact center of the circle.
4. Tighten the knurled wheels in the direction of the
arrow.
The foot screws are secured against being
adjusted.
ATTENTION!
Incorrect measurement due to force shunts
If the scale and housing is in contact with some-
thing, e.g. a towel, weight will not be measured
correctly.
– Set the scale up so that only its foot screws
are in contact with the floor.