Packaging for Sterilization

The clean, dry instruments should be placed in the Steribox sterilization container. The container should
be bagged in a sterilization pouch that is validated for use in a 132ºC pre-vacuum steam sterilization
cycle. The sterilization pouch should be sealed according to the manufacturer's instructions.

Sterilization

The reusable Hu-Friedy & EMS Piezon 150 and 250 handpieces, instruments, and accessories must
be steam-sterilized prior to the first use and following each patient use. Hu-Friedy & EMS recommend
sterilization in a fractional prevacuum cycle at 132°C for 4 minutes with a drying time of 20 minutes.
Hu-Friedy & EMS recommend use of biological indicators and chemical indicators that have been validated
for use in a 132ºC gravity-displacement steam sterilization cycle.