The measured value of the BENNING MM 8 is obtained by a mean value recti-
fication and displayed as an effective value.
The measured value of the BENNING MM 9/ 10 is obtained as a real effective
value (­TRUE RMS) and then displayed.
The measuring accuracy is specified for sinusoidal curves and is valid for
*1
indicating values below 4000 digits. For indicating values higher than 4000
digits 0.6 % have to be added to the specified measuring accuracy. In case
of non-sinusoidal curves below 2000 digits, the indicating value becomes
inaccurate. Thus, an additional error occurs for the following crest factors:
crest factor from 1.4 to 3.0 additional error ± 1.5 %
valid for sinusoidal curves 50 Hz/ 60 Hz
*2
7.3­­Direct­current­ranges
overload protection:
-
600 V
at the µA input,
eff
-
10 A (­500 V) fuse, quick-acting, at the 10 A input (­BENNING MM 9/ 10),
Measuring
Resolution
range
600 µA
6000 µA
6 A
10 A
7.4­ Alternating­current­ranges­(­BENNING MM 9/ 10)
overload protection:
-
10 A (­500 V) fuse, quick-acting, at the 10 A input (­BENNING MM 9/ 10),
Measuring
Resolution
range
6 A
10 A
The measured value of the BENNING MM 8 is obtained by a mean value recti-
fication and displayed as an effective value.
The measured value of the BENNING MM 9/ 10 is obtained as a real effective
value (­TRUE RMS) and then displayed.
The measuring accuracy is specified for sinusoidal curves and is valid for
*1
indicating values below 4000 digits. For indicating values higher than 4000
digits 0.6 % have to be added to the specified measuring accuracy. In case
of non-sinusoidal curves below 2000 digits, the indicating value becomes
inaccurate. Thus, an additional error occurs for the following crest factors:
crest factor from 1.4 to 3.0 additional error ± 1.5 %
7.5­ Resistance­measuring­ranges
Overload protection in the case of resistance measurements: 600 V
Measuring
Resolution
range
600 Ω
6 kΩ
60 kΩ
600 kΩ
6 MΩ
60 MΩ
7.6­ Diode­and­continuity­testing
The indicated measurement precision applies in the range between 0.4 V and
0.8 V.
Overload protection for diode testing: 600 V
The built-in buzzer sounds in the case of a resistance R less than 100 Ω.
Meas.
resolu-
range
tion
10 mV
11/ 2006
0,1 µA
± (1.0 % of the measured value + 2 digits)
1 µA
± (1.0 % of the measured value + 2 digits)
1 mA
± (1.0 % of the measured value + 2 digits)
10 mA
± (1.0 % of the measured value + 2 digits)
Meas. precision
frequency range 50 Hz - 500 Hz
1 mA
± (1.5 % of the measured value + 5 digits)
10 mA
± (1.5 % of the measured value + 5 digits)
0,1 Ω
± (0.7 % of the measured value + 2 digits)
1 Ω
± (0.7 % of the measured value + 2 digits)
10 Ω
± (0.7 % of the measured value + 2 digits)
100 Ω
± (0.7 % of the measured value + 2 digits)
1 kΩ
± (0.7 % of the measured value + 2 digits)
10 kΩ
± (0.7 % of the measured value + 2 digits)
Meas. precision
± (1.5 % of the measured value + 5 digit)
­ B ENNING­MM­8/­9/­10­
Meas. precision
*1
within the
Meas. precision
eff
Voltage drop
< 4 mV/ µA
< 4 mV/ µA
2 V max.
2 V max.
Voltage drop
2 V max.
2 V max.
eff
Max. idling
voltage
1,3 V
1,3 V
1,3 V
1,3 V
1,3 V
1,3 V
Max. meas.
Max. idling
current
voltage
1,5 mA
3,0 V
5