Operating Instructions Midwest E Pro / E Plus
66 04 206 D3686.201.01.02.02
Article
Holding time at
135 °C (275° F)
Wrapped instruments 10 minutes
Unpackaged
3.5 minutes
instruments
Gravity displacement steam sterilizers are permitted.
without DAC PROFESSIONAL _ US
NOTICE! Do not exceed 140°C (284°F), even during the drying phase.
After sterilizing
After sterilizing - heading
1. Remove the instrument from the steam sterilizer immediately.
Subsequent steps_instruments
CAUTION! The instrument is hot. Risk of burns!
NOTICE! Do not attempt to accelerate the cooling process by
immersing the instrument in cold water. This can damage your
instrument.
2. Store all instruments so that they are protected from contamination.
3. Sterilize again once the storage period has elapsed.
7 Conditioning
Drying time
30 minutes
0 - 1 minute
49