7 Conditioning
46
✔ The O-ring on the spray adapter is intact.
1. Fit the spray adapter onto the nozzle of the spray can.
2. Insert the instrument until it snaps into place and hold it.
3. Spray the instrument for 1 - 2 seconds.
IMPORTANT: Hold the spray can upright.
4. Wipe any spray that comes out with a disinfection cloth.
5. Repeat the process until the spray is clear.
7.4.2
Care of the push button chuck
Use Midwest® Plus Aerosol Spray on the push button chuck to remove
New
Sirona
deposits and ensure proper functioning of the clamping system.
Interval
Interval
● At least once a week
Operating Instructions Midwest E Pro / E Plus