7 Conditioning
48
7.5
Sterilizing
Intervals:
Interval _ US
● Prior to initial operation
● Prior to every other use
Process
✔ The instrument is clean and disinfected.
Action steps_instruments
✔ Instrument maintenance is complete.
✔ The instrument can be sterilized in packaging suitable for
sterilization and storage: paper/plastic composite packaging
according to ISO 11607.
➢ Sterilize the instrument in the steam sterilizer with saturated water
vapor.
Unit pressure in bar and psi_US
Temperature:
135° C (275° F)
Overpressure:
2.13 bar (30.89 psi)
Operating Instructions Midwest E Pro / E Plus