Télécharger Imprimer la page

Dentsply Sirona Midwest E Pro Notice D'utilisation page 14

Masquer les pouces Voir aussi pour Midwest E Pro:

Publicité

Les langues disponibles

Les langues disponibles

6604206.book Seite 12 Mittwoch, 30. November 2016 5:54 17
2 Safety information
Obligations of the user
Preventing the spread of infections and cross
contamination
Emitted cooling air
12
2
Safety information
● Use only fault-free materials that do not deviate from the specified
Obligations of the user
data [ → 19].
● Protect yourself, patients, and others against any foreseeable
dangers. To do this, follow the safety information.
● Comply with the Intended use of the equipment.
● You should always keep these operating instructions within reach for
further reference.
Prevent the spread of infections and cross contamination between
Sterilization
Cross contamination and spread of infections
patients, users, and third parties. Sterilize equipment after each patient.
Take the appropriate hygiene measures, e.g. wear protective gloves.
The cooling air emitted by the coupling of the motor must have a flow
Emitted cooling air
rate of 1.5 - 10 Nl/min.
To keep the cooling air coming out of the instrument away from the
Keeping cooling air away
cavity, silicone disks are available [ → 37].
Operating Instructions Midwest E Pro / E Plus
66 04 206 D3686.201.01.02.02

Publicité

loading

Ce manuel est également adapté pour:

Midwest e plus