Operating Instructions Midwest E Pro / E Plus
66 04 206 D3686.201.01.02.02
7.2
Automated cleaning and disinfecting ...
7.2.1
... with cleaning and disinfection equipment
The instrument can also be cleaned and disinfected in a suitable piece
Instrument
of cleaning and disinfection equipment.
The cleaning and disinfection equipment used must be approved by its
Cleaning and disinfection equipment
manufacturer for the cleaning and disinfection of dental instruments and
comply with EN ISO 15883-1 (e.g. 95°C (203°F) and 10 min. holding
time).
For further details, refer to the operating instructions supplied with the
unit.
✔ The instrument is conditioned with a cleaning and disinfection
Kontrolle _ Instrument
device.
1. Check whether the instrument is clean after conditioning under good
lighting (min. 500 lux) and color rendering index (min. 80 Ra).
2. If it is still dirty, repeat the process.
7 Conditioning
41