Required accessories
T1 spray
A
Black nozzle insert
B
Midwest® Plus Aerosol Spray
NOTICE! Only use Midwest® Plus Aerosol Spray.
Using the T1 spray
Procedure
✔ The instrument is clean and disinfected.
Vorgehensweise _ Winkelstück
1. Press the contra-angle handpiece head with the chuck firmly against
the spray can nozzle.
2. Spray the chuck for 1 - 2 seconds.
IMPORTANT: Hold the spray can upright.
3. Wipe any spray that comes out with a disinfection cloth.
7 Conditioning