8  M aintenance
8.1  M eter maintenance
Never unscrew the two halves of the housing!
The meters do not require any maintenance other than an occasional wipe with a damp cloth. The
hous i ng is made of acrylonitrile butadiene styrene/polycarbonate (ABS/PC). This material is sen s i-
tive to some organic solvents, such as toluene, xylene and methyl ethyl ketone (MEK).
Any spillage should be wiped off immediately.
8.2  D isposal
In compliance with European Directive 2002/96/EC on Waste Electrical
and Elec t ron i c Equipment (WEEE), this instrument must not be disposed of
togeth e r with domestic waste. This also applies to countries outside the EU,
per their specific requirements.
Please dispose of this product in accordance with local regulations at the
col l ect i ng point specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the
dis t rib u t or from which you purchased this instrument.
Should this instrument be passed on to other parties (for private or pro f es-
sion a l use), the content of this regulation must also be related.
Thank you for your contribution to environmental protection.
8.3  E rror messages
Message
Con d uc t iv i t y/TDS/salin i t y/resis t iv i t y/
con d uc t iv i t y ash/temperature exceeds
max. limit
Con d uc t iv i t y/TDS/salin i t y/resis t iv i t y/
con d uc t iv i t y ash/temperature below
min. limit
Memory is full
Please calibrate electrode
Description and Resolution
Measurement limits are activated in the menu settings
and measured value is outside these limits.
Check the sample.
Check sample temperature.
Make sure that the pH electrode wetting cap has
been removed and that the electrode is properly con-
nect e d and placed in the sample solution.
Max. 1000 measurement data can be stored in the
mem o r y.
Delete all or partial data in the memory, otherwise
you will not be able to store new measurement data.
Calibration reminder has been switched on in the menu
set t ings and last calibration has expired.
Calibrate the electrode.

Maintenance