Message
Wrong password
Passwords do not match
Program memory error
Data memory error
No matching data found in memory
Sensor ID already exists, previous SN
will be overwritten
Standard temp out of range
Temp. out of nLF correction range
Temp. out of pure water range
Temp. out of conductivity ash cor r ec-
tion range
Update failed
Description and Resolution
The entered PIN is not correct.
Re-enter the PIN.
Reset to factory settings, all data and settings will
be lost.
The confirmation PIN does not match with the entered
PIN.
Reenter PIN.
Meter recognizes internal error during start-up.
Switch the meter off and back on.
Contact METTLER TOLEDO service if the problem per-
sists.
The data could not be stored into memory.
Switch the meter off and back on.
Contact METTLER TOLEDO service if the problem per-
sists.
The entered filter criterion does not exist.
Enter a new filter criterion.
Two sensors with the same ID but different SN are not al-
lowed in the meter. If a different SN has been entered for
this sensor ID previously, the old SN will be over w rit t en.
Enter a different Sensor ID in order to keep the pre v i-
ous ID and SN.
Conductivity calibrations can only be performed at tem-
per a t ures from 0 ... 35°C.
Keep the standard temperature within the range.
Conductivity measurements of natural water can only be
per f ormed at temperatures from 0 ... 36°C.
Keep the sample temperature within the range.
Conductivity measurements of pure water can only be
per f ormed at temperatures from 0 ... 50°C.
Keep the sample temperature within the range.
Conductivity ash measurements can only be performed
at temperatures from 15 ... 25°C.
Keep the sample temperature within the range.
The software update process failed. This could be due to
the following reasons:
The USB stick is not connected or it is disconnected
dur i ng the update process
The update software is not in the correct folder