5. How long will the tip last?
It depends on how well you take care of the tip. We have
had customers using the same tip for two years while other
customers will use up a tip in two weeks. The life of a tip will
be reduced by:
a. running it too hot for long periods of time
b. bending or putting too much mechanical force on it.
The tips should be used only to transfer heat—not to pack
down gutta-percha. Our recommendation is that you use a cold
plugger for that purpose. Tip life will also be greatly extended
if you use the touch mode consistently so that the tip is heated
only when needed.
6. How does the touch switch work and why do I need a
contact clip?
A spring extending backward from the pinvise mount
completes the circuit by making contact with the probe shaft.
7. What kind of batteries are used?
The Touch 'n Heat uses sealed lead-acid batteries. Unlike the
common nickel-cadmium batteries, you should never run down
the lead-acid batteries completely. Total depletion will destroy
them. Our Touch 'n Heats have low battery warning lights to
prevent this from happening. Always keep the unit on charge
when not in use.
8. How long do the batteries hold a charge?
The Touch 'n Heat battery will last 30–45 minutes and can be
recharged in 8 hours. For your safety, the Touch 'n Heat is
designed not to function while being charged.
9. When will the batteries have to be replaced?
The battery manufacturer specifies an estimated battery life of
3 to 5 years when kept properly charged. When the batteries are
depleted, they must be removed and replaced by SybronEndo
since the batteries are soldered directly into the unit.
10. How should I use the Touch 'n Heat?
Practitioners who ask this general question probably do not
know enough about the warm gutta-percha technique. We
usually recommend attending some courses on this subject.
You should not start using the Touch 'n Heat without having
some course instruction and prior practice on extracted teeth.
Cleaning, Sterilization, and Maintenance
Sterilization
Pluggers and tips must be cleaned and sterilized before every use:
1. Clean with water and mild, non-abrasive, detergent such as dish
washing liquid. Dry thoroughly.
2. Double wrap the pluggers and tips using a typical central
service wrapping technique and place into autoclave unit
avoiding contact with other instruments.
3. Operate sterilization cycle at 250°F for 30 minutes or 270°F
for 25 minutes. Follow the instructions supplied with your
autoclave unit for specific operating instructions.
Expected Life of Accessories
All accessories are reusable only for a limited number of
procedures. This number is highly dependent on how and for
what duration the instrument is used during each procedure and,
therefore, will vary significantly from user to user. If the device
function appears to be erratic, an accessory may have exceeded
its useful life and should be replaced.
Cleaning
The Touch 'n Heat 5004 enclosure and the handpiece should
be wiped down with a cloth soaked in any of the common
non-abrasive, mild, cleaning liquids containing alcohol. Do
not submerge the device or let liquid enter the enclosure
and handpiece.
Technical Speci cations
Dimension/Weight
Casing of unit:
Weight: 1.4 lb. (0.64 kg)
Dimensions: Height: 2.2 inches /55 mm
Classification: IEC 601-1 Internally Powered Equipment
Mode of Operation: IEC 601-1 Continuous Operation
Duty Cycle: 2 sec. ON, 5 sec. OFF for 5 mins.
@ 30 mins. Interval
Degree of safety of application in presence of a flammable
anesthetic mixture:
Width: 3.1 inches /79 mm
Depth: 5.2 inches /132 mm
3
Unit unsuitable for use in presence of a flammable anesthetic
mixture with air or with oxygen or nitrous oxide.
Conditions for use: +10°C to +40°C
30% to 75% relative humidity
Storage and transport conditions: -10°C to +45°C
Atmospheric Pressure: 101.1 kPa
Origin: SybronEndo
1332 South Lone Hill Avenue
Glendora, CA 91740, USA
Model: Touch 'n Heat 5004
Power Supply: rechargable sealed lead-acid battery
Warranty Information
Returns and Warranty
The Touch 'n Heat is designed for lasting durability and predictable
results. The unit comes with a one-year warranty.
If the unit should fail to operate correctly, please contact your
SybronEndo customer service representative at 1 800-346-3636.
Outside the United States, please call customer service at
714-516-7979 or contact your local dealer or supplier to
coordinate the returns/repairs with SybronEndo.
Warranty repairs can only be performed by SybronEndo or by
authorized agencies with original factory parts. Any unauthorized
repairs will void the warranty.
1. Use the original packaging when shipping or storing your
Touch 'n Heat.
2. Many problems can be solved over the phone. Do not hesitate
to contact us if you experience difficulties when using the
Touch 'n Heat.
3. For returns, call us for a Return Good Authorization (RGA)
number.
4. Damages incurred during shipping due to improper packaging
are not covered by the warranty. If the original box and/or foam
packing are not available, please call SybronEndo for packing
instructions.
5. Mark all outer packaging with the RGA number, your name,
address, and phone number.
6. Shipping costs are not covered by the warranty.
Disclaimer
The responsibility of SybronEndo, as a manufacturer of
electrical/medical devices, extends to the technical safety features
of the device only if maintenance and repairs are carried out by
SybronEndo or by authorized agencies with original factory parts.
For safety reasons, this product should be used with accessories
manufactured and sold by SybronEndo. Any use of non-authorized
accessories or not following any of the instructions for use is done
so at the operator's risk and voids the warranty.
SybronEndo does not assume any responsibility for poor results
due to operator error or equipment malfunction.
10% to 95% relative humidity