Adjusting Transmitted-Light Brightfield On The Fixed-Köhler Microscope - Zeiss Primo Star Mode D'emploi

Masquer les pouces Voir aussi pour Primo Star:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 79
Primo Star / Primo Star iLED
• Turn the knurled knob for vertical adjustment of the condenser (Fig. 16/22) until the edge of the
luminous-field diaphragm appears sufficiently sharp (Fig. 21/B).
• Center the image of the luminous-field diaphragm using both centering screws (Fig. 21/5) of the
condenser (Fig. 21/C). Then, open the diaphragm until it just disappears from the field of view
(Fig. 21/D).
• To adjust the aperture diaphragm (contrast), remove one eyepiece from the tube and look through the
tube with your naked eye. Swing lever (Fig. 21/4) to adjust the aperture diaphragm to approximately
2/3 ... 4/5 of the diameter of the exit pupil of the objective (Fig. 21/E). In most applications, this
aperture diaphragm setting provides optimum contrast at almost ideal resolution, and is therefore the
best compromise for the human eye.
• Insert the eyepiece again into the tube.
The size of the field of view and the objective aperture change after every objective change.
Therefore, repeat the adjustment of luminous-field diaphragm and aperture diaphragm to
obtain optimum results.
3.4.4
Adjusting transmitted-light brightfield on the Fixed-Köhler microscope
The Primo Star Fixed-Köhler microscope is supplied factory-adjusted. Operation is restricted to a few
manipulations.
• Place the specimen in the specimen holder of the mechanical stage.
• If the microscope stand is equipped with a dark-field slider, pull this slider to the left up to the lock-in
position.
• When using the Primo Star iLED, turn the transmitted light / reflected light changeover switch to the
transmitted light position (Brightfield) (turn it first fully upward). Push the slider containing the yellow
filter with its filter position into the light path.
• Adjust the desired magnification by rotating the corresponding objective into the light path.
• Set the control lever of the condenser aperture diaphragm to the value of the selected magnification
(10x, 40x or 100x).
• Focus onto the specimen using the focusing drive.
• Turn the rotary knob on the microscope stand to adjust the illumination intensity to a level pleasant
for observation.
If the condenser had been removed (e.g. for installing the illuminating mirror), make sure to
reinstall and center it by means of the two centering screws (for this, refer to section 3.5.5).
M60-2-0011 v 08/2008
Start-Up and Operation
Carl Zeiss
27

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Primo star iled

Table des Matières