Cleaning set CLO/DGMa for sensor types CLO
1
Cleaning set CLO/DGMa for sensor types CLO
User qualification: instructed user
n
Only use the cleaning set CLO/DGMa (1104286) for the hydrodynamic cleaning of sensor
types CLO in conjunction with the sensor fitting DGMa.
Increased sensitivity (slope)
The sensitivity (slope) of the sensor increases by approx. 20% with hydrodynamic cleaning,
with the upper measuring range limit falling correspondingly with types CLOx-ma.
The optional cleaning set CLO/DGMa no. 1104286 is recommended with applications that lead to
slowly accumulating deposits (e.g. limescale, manganese or iron oxides, or biofilm) on the surfaces
of the electrodes.
Only use the cleaning set with the bypass fitting DGMa.
Tab. 1: Cleaning set, part number 1104286
Parts
Flow nozzle CLO
Set of cleaning balls
The electrodes are cleaned by the cleaning balls continuously hitting the surfaces of the electrodes
caused by the flow of water. The cleaning balls remove the deposits, preventing the formation of the
deposits.
4
Quantity
1
approx. 100
Part number
1104264
1104267