Technical Data; Dimensions And Weight; Electrical System; Environmental Conditions - Invacare Robin Manuel D'utilisation

Masquer les pouces Voir aussi pour Robin:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 47
Invacare® Robin®

10 Technical Data

10.1 Dimensions and weight

Dimensions
Hoist length (A)
Hoist width (B)
Hoist height (C)
Maximum lifting range (D)
Weights
Maximum lifting capacity (Safe
Working Load)
Total weight without sling

10.2 Electrical System

Maximum current
input
Protection class
1
(ceiling hoist: IP24; hand control: IP44)
Insulation class
Applied part complying with the
specified requirements for protection
against electrical shock according
Sound level
Number of lifts
65 lifts of 0.5 m with 80 kg
per charge
20
Robin®
Robin®
Mover
470 mm
470 mm
250 mm
250 mm
203 mm
266 mm
2.5 m
2.5 m
Robin®
Robin®
Mover
200 kg
200 kg
7.3 kg
10 kg
Robin®
Robin® Mover
0.9 A
Entire device: IPx4
Class II equipment
Type B applied part
to IEC60601-1.
50 – 55 dB (A)
Intermittent
(periodic motor
10%, max. 2 min. / 18 min.
operation)
Battery
NiMH Battery 24 V / 1.9 Ah
Charging time
Lifting speed with
2.1 cm/s
200 kg load
Traverse speed in
the tracks
Charger
During charging: 25 – 36 V DC
Voltage output
100 – 240 V AC, 50 – 60 Hz
Voltage supply
Protection class
1
IP41 or IP67 (depending on version)
1
See product label and label on each electric device for
correct protection class. The lowest IP classification decides
the overall classification of the device.
IPx4: Protected against water splashed from any
direction.
IP24: Protected against objects larger than 12.5 mm and
protected against water splashed from any direction.
IP44: Protected against objects larger than 1 mm and
protected against water splashed from any direction.
IP41: Protected against objects larger than 1 mm and
protected against vertically dripping water.
IP67: Dust-tight and protected against temporary
immersion into water.

10.3 Environmental conditions

Storage and
transportation
Temperature
-30 °C to +50 °C
Relative humidity
10 % to 70 %
Atmospheric
pressure
Allow the product to reach operation temperature
before usage:
Warming up from minimum storage temperature
might take up to 24 hours.
Cooling down from maximum storage
temperature might take up to 24 hours.
Robin®
Robin® Mover
Up to 3 hours at 20°C
2.1 cm/s
n/a
14 cm/s
No-load: 41 V DC
Operation
+5 °C to +40 °C
20 % to 90 % at 30
°C, not condensing
700 hPa to 1060 hPa
1586745-E

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Robin mover

Table des Matières