10.6 Electromagnetic compliance (EMC)

Guidance and manufacturer´s declaration – electromagnetic emission
The patient lift is intended for use in the electromagnetic environment specified below. The customer or the user of the patient
lift should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group I
CISPR 11(partly)
RF emissions
Class B
CISPR 11(partly)
Harmonic
emissions
Class A
IEC 61000-3-2
Voltage fluctuations
/ flicker emissions
Complies
IEC 61000-3-3
Guidance and manufacturer´s declaration – electromagnetic immunity
The patient lift is intended for use in the electromagnetic environment specified below. The customer or the user of the patient
lift should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
1575480-E
Electromagnetic environment - guidance
The patient lift uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
The patient lift is suitable for use in all establishments including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Compliance level
± 6 kV contact
± 8 kV air
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Technical Data
47