Invacare® Birdie®
Checking the boom pin
1. Turn carabiner B to
the side.
2. Remove plastic part A.
3. Check the welded pin
C for wear.
Do not use the lift if
the pin measures less
than 7 mm.
4. Insert the plastic part
again and release the
carabiner to its position.
7.6 Replacing the Hanger Bar
If the scale is installed on the lift, refer to the scale
instruction manual to replace the hanger bar.
1. Support the hanger bar A with one hand.
2. Open the carabiner B by first twisting the safety catch
and then press it backwards with one hand.
40
3. Remove the hanger bar from the carabiner.
4. Reverse STEPS 1- 3 to install the new hanger bar.
5. Check that the hanger bar is firmly attached to the
carabiner. The gate of the carabiner must be closed
after the hanger bar is mounted.
1575480-E