OPERATION
OPERATING CONTROLS  TREATMENT SCREEN CONTINUED
(11) Speed
The percentage of speed of when
the Traction Cord is being pulled or
released. There are three speed options
to choose from: 30%, 50% or 100%.
(12) Pre-tension
Pre-tension allows the clinician to
set a certain level in weight and time
to allow the opportunity to make
adjustments to harnesses before
treatment starts.
(13) Edit Hold/Rest Times
Displays the amount of hold and rest
times (0-99 seconds) of Progressive,
TX and Regressive traction entered
during setup. To enter the hold or rest
time, press the appropriate button,
then enter the desired time, using
the number keypad. Press the Return
Arrow button to accept.
NOTE: The Hold and Rest Time can be
changed during the traction
therapy session by pressing the
35
Triton® Traction
Edit Hold/Rest Times button,
press the appropriate button,
and use the Up or Down Arrow
buttons any changes. Press the
Return Arrow button to accept.
(14) Edit Min/Max Levels
Displays the Minimum and Maximum
Traction Tension entered during
setup. To enter the Minimum or
Maximum Traction Tension, press the
Edit Min/Max Levels button, press the
appropriate button, then enter the
desired setting, using the number
keypad. Press the Return Arrow button
to accept.
NOTE: The Minimum and Maximum
Traction Tension can be
changed during the traction
therapy session by pressing the
Edit Min/Max Levels button,
press the appropriate button,
then use the Up or Down
Arrow buttons to make
changes in 1 lb (1 kg or 1 N)
increments. Press the Return
Arrow button to accept.