SPECIFICATIONS
ELECTROMAGNETIC COMPATIBILITY TABLES CONTINUED
The Triton traction unit is intended for use in the electromagnetic environment specified in the table below. The customer or user of the Triton traction unit should assure
that it is used in such an environment.
Immunity Test
Immunity Test
IEC 60601
Test Level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power supply lines
transient/burst
±1 kV for input/output lines
IEC 61000-4-4
Surge
±1 kV differential mode
IEC 61000-4-5
±2 kV common mode
Voltage dips, short
<5% U T
interruptions and
(>95% dip in U T ) for 0.5 cycle
voltage variations
40% U T
on power supply
(60% dip in U T ) for 5 cycles
input lines
70% U T
IEC 61000-4-11
(30% dip in U T ) for 25 cycles
<5% U T
(>95% dip in U T ) for 5 sec
Power frequency
3 A/m
(50/60Hz)
magnetic field
IEC 61000-4-8
NOTE:
U T is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration - electromagnetic immunity
Compliance Level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
<5% U T
(>95% dip in U T ) for 0.5 cycle
40% U T
(60% dip in U T ) for 5 cycles
70% U T
(30% dip in U T ) for 25 cycles
<5% U T
(>95% dip in U T ) for 5 sec
3 A/m
Electromagnetic Environment - Guidance
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity
should be at least 30%
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Triton traction
unit requires continued operation during power mains interruptions, it is recommended that the Triton traction unit
be powered from an uninterrupted power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or
hospital environment.
18
Triton® Traction