Triton 4739 Mode D'emploi page 22

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

SPECIFICATIONS
ELECTROMAGNETIC COMPATIBILITY TABLES CONTINUED
The Triton traction unit is intended for use in the electromagnetic environment specified in the table below. The customer or user of the Triton traction unit should assure
that it is used in such an environment.
Immunity Test
Immunity Test
IEC 60601
Test Level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
3 V/m
Radiated RF
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the Triton traction is used exceeds the applicable RF compliance level above, the Triton
traction should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the
Triton traction.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V
Guidance and manufacturer's declaration - electromagnetic immunity
Compliance
Level
Portable and mobile RF communications equipment should be used no closer to any part of the Trition traction unit, including cables, than
the recommended separation distance calculated from the equation applicable to the frequency of the transmitter
Recommended separation distance
d = [3.5]√P
3 V
V
1
d = [3.5]√P 80 MHz to 800 MHz
E
1
3 V/m
d = [7]√P 800 MHz to 2.5 GHz
E
1
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey
b
each frequency range
.
Interference may occur in the vicinity of equipment marked with the following symbol:
Electromagnetic Environment - Guidance
1 ] V/m.
19
Triton® Traction
a
, should be less than the compliance level in

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

47782841

Table des Matières