How the keys work
NOTICE
A program number is given to each of the different modes as shown in the table below. Only the preprogrammed
modes can be selected.
Program No
1
2
3
SURG
4
5
6
7
8
9
*1 Burst function: Vibration levels can be changed at certain intervals. Levels can be selected depending on patient's bone
hardness (density).
*2 Amount of water may vary slightly depending on the condition of the Irrigation Tube.
Select Mode Key
For switching to SURG, ENDO, and PERIO modes. Switching modes will automatically change the program number and
power range.
Program Key
For selecting a program number. Press the "+" key to increase the number by one, or the "-" key to decrease by one.
The program numbers that can be selected are different depending on the mode selected. Program numbers can't be
changed while the Handpiece is in operation.
Power Key
For adjusting power level. Press the "+" key to increase the number by one, or the "-" key to decrease by one. The power
range is different depending on the mode selected. This key can be used while the Handpiece is in operation.
Flow Key
For adjusting the flow level. There are 5 flow levels to choose from (see the table above).
Auto Cleaning mode
Pressing the Flow Key for 3 seconds will activate the Auto Cleaning mode in which the pump rotates at maximum
flow rate for 30 seconds to clean the inside of handpiece. Refer to 8-1 Auto Cleaning for more details.
MEMORIZE Key
For saving settings. Press the key for about 1 second, and the current settings will be saved under the program number
selected and a beep will let you know memorizing is complete.
CAUTION
Be careful not to place heavy things on the LCD display of the Operation Panel and avoid any strong impact (e.g.
dropping something onto it). This may lead to damage or malfunction.
Vibration mode
1
With Burst function *
10Hz
1
With Burst function *
30Hz
1
With Burst function *
60Hz
1
No Burst function *
1
No Burst function *
ENDO
PERIO
Power range
10-100%, in 10%
increments
5-50%, in 5% increments
60% and up, in 10%
increments
8
Flow level (5 levels)
2
15~90ml/min*
2
9~38ml/min*