Télécharger Imprimer la page

NSK VarioSurg Mode D'emploi page 4

Publicité

Les langues disponibles
  • FR

Les langues disponibles

•Only Use Tips that NSK has approved for VarioSurg use. Damage, failure or an accident may occur; the following are
possible failure modes (NSK Warranty Exclusions).
• Vibration failure caused by using non conforming screws (Incorrect Tip).
• Patients' accidental ingestion of Broken or damaged Tip.
• Damage to thread of Handpiece output shaft.
• Handpiece, Handpiece Cord, Tips, Tip Wrench, and Tips Holder are all non-sterile items. Be sure to sterilize them
before use.
• Never sterilize the Handpiece etc., using ultraviolet sterilization. It may cause discoloring.
•After autoclave sterilization, wipe away any remaining water drops. They may cause discoloring.
•Do not press the Tip against anything unless it is part of treatment. Ultrasonic vibrations may damage tooth surface.
•The Tip wears down after every use. After wearing down, vibrations become weak and coating may come off. If these
problems occur, be sure to replace the Tip with a new one. It is recommended that the Tip be changed after 5 uses.
•When replacing the Tip, use the Tip Wrench to ensure that the new Tip is securely mounted to the Handpiece. If
mounting is loose, vibrations may be weak and output inadequate.
•If a new Tip is mounted with stray objects on the screw, vibrations may be weak and output inadequate. If there is
anything on the Tip, clean it off before mounting to Handpiece.
•If it feels as though the Tip is not vibrating during use, remove the Handpiece from the mouth of the patient, and step
on the foot control again. If the situation is no better, the Tip may have become loose during use, so check to make
sure it is securely mounted.
• Do not shine the LED into the eyes of the patient or the operator. It may damage eyesight.
• This product is Medical Electrical equipment. EMC (Electromagnetic compatibility) is described in the accompanying
documentation.
•This product needs special precautions regarding EMC and needs to be installed and put into service according to
the EMC information.
•Portable and mobile RF communications equipment can affect Medical Electrical equipment. Do not use RF
equipment outskirts for the product.
• The use of ACCESSORIES, transducers and cables other than those specified, with the exception of transducers
and cables sold by the manufacturer of this product as replacement parts for internal components, may result in
increased EMISSIONS or decreased IMMUNITY of this product.
•This system should not be used adjacent to or stacked with other
equipment and that if adjacent or stacked use is necessary, this product
should be observed to verify normal operation in the configuration in
which it will be used.
•Grounding reliability can only be achieved when the equipment is
connected to an equipment receptacle marked "Hospital Only" or
"Hospital Grade".
• During vibration, the Handpiece and the Handpiece Cord may affect computer and LAN cable. Noise could be heard
during operation near a radio receiver.
• Turn the Main Power Switch OFF when not in use. If left unused for a long period of time, remove the plug from the
power outlet.
• Clean and sterilize immediately after use and before storing. Lack of cleaning or sterilization could result in
breakdowns.
• When you have not used the product for long time and use it again, check the operation before use.
• Users are responsible for operational control, maintenance and inspection.
• No special training is required for this device.
• Applied parts for patient and/or operator are/is Tip and Handpiece.
CAUTION
NOTICE
3

Publicité

loading

Ce manuel est également adapté pour:

Ne214