Recommended separation distances between portable and mobile RF communications equipment and the VarioSurg
The VarioSurg is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the VarioSurg can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the VarioSurg as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Cables and accessories
Maximum length
Handpiece cord
2m
Foot Controller
2m
Separation distance according to frequency of transmitter
150kHz to 80MHz
d = 1.2 P
0.12
0.38
1.2
3.8
12
Complies with
RF emissions, CISPR11, EN55011
Harmonic emissions,
Voltage fluctuations/ flicker emission,
Electrostatic discharge(ESD),
Electric fast transient / burst
Surge
Voltage dips, short interruptions and voltage variations on power supply input lines
Power frequency(50/60Hz) magnetic field
Conducted RF
Radiated RF
24
m
80MHz to 800MHz
d = 1.2 P
0.12
0.38
1.2
3.8
12
800MHz to 2.5GHz
d = 2.3 P
0.23
0.73
2.3
7.3
23
Class B/ Group 1
EN/IEC61000-3-2
EN/IEC61000-3-3
EN/IEC61000-4-2
EN/IEC61000-4-4
EN/IEC61000-4-5
EN/IEC61000-4-11
EN/IEC61000-4-8
EN/IEC61000-4-6
EN/IEC61000-4-3