► Check the product for visible damage before each use.
CAUTION
Independent manipulation of the product
Injury due to product damage or malfunction.
► Manipulations to the product other than the tasks described in these
instructions for use are not permitted.
► The product and any damaged components may only be opened and
repaired by authorised, qualified Ottobock personnel.
CAUTION
Distance to HF communication devices is too small (e.g. mobile
phones, Bluetooth devices, WiFi devices)
Injury due to unexpected behaviour of the product caused by interference
with internal data communication.
► Therefore, keeping the following minimum distances to these HF com­
munication devices is recommended:
• Mobile phone GSM 850/GSM 900: 0.99 m
• Mobile phone GSM 1800/GSM 1900/UMTS: 0.7 m
• DECT cordless phones incl. base station: 0.35 m
• WiFi (routers, access points,...): 0.22 m
• Bluetooth devices (third-party products not approved by Ottobock):
0.22 m
NOTICE
Improper product care
Damage to the product due to the use of incorrect cleaning agents.
► Only clean the product with a damp cloth and mild soap (e.g.
453H10=1 Ottobock DermaClean).
► Do not bring the surface of the electrodes into contact with a special
cleaner for Ottobock cosmetic gloves, e.g 640F12.
5 Scope of Delivery and Accessories
5.1 Scope of delivery
1 pc. 13E202=50/13E202=60 suction socket electrode
1 pc. 13E80 adjustment tool
1 pc. Instructions for use (qualified personnel)