4 Scope of delivery
Quantity
1
1
Spare parts/options (not included in the scope of delivery)
Designation
Connection plate
Only for 1D10 without adapter and for 1D11
Foot adapter with screw connection
Ankle block (exoskeletal design)
Screw connection for ankle block (exo­
skeletal design)
Ottobock special adhesive with hardener
5 Preparing the product for use
CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
► Observe the alignment and assembly instructions.
NOTICE
Grinding the prosthetic foot
Premature wear resulting from damage to the prosthetic foot
► Do not grind the prosthetic foot.
5.1 Installing the foot adapter or ankle block
The installation must be performed if the prosthetic foot is not
equipped with a foot adapter or ankle block.
Designation

Instructions for use

Prosthetic foot
Reference number
2R14
2R8*, 2R31*, 2R54*
2K34=*
2Z22=M8x90,
2Z22=M10
636W28 or 636W18
and 636W19
5.1.1 Installing the foot adapter
► Install the foot adapter on the prosthetic foot according to its
instructions for use.
5.1.2 Installing the ankle block (exoskeletal design)
Required materials : Screw connection (see Page 11), 710D4
>
torque wrench, 40 grit sandpaper, 617H21 sealing resin, 617P37
hardener powder, suitable solvent (e.g. 634A28), only for
1WR95: 636W17 plastic adhesive
1) Insert the threaded socket of the screw connection in the proximal
side of the hole in the ankle block.
2) Insert the screw on the foot sole side in the prosthetic foot and
screw the ankle block to the prosthetic foot (mounting torque:
30 Nm).
3) Conduct the fitting of the ankle block.
4) Loosen the screw connection between the prosthetic foot and the
ankle block.
5) Sand the ankle block according to the contour of the prosthetic
foot.
6) Roughen the connection surface of the prosthetic foot in an a – p
direction using the sandpaper.
7) Remove the dust from the sanded surfaces.
8) Mix the sealing resin with 2% to 3% hardener powder for the con­
nection between the ankle block and the prosthetic foot.
9) INFORMATION: Do not put the resin in the hole for the
screw connection.
Apply the resin evenly to the connection surfaces of the ankle
block and prosthetic foot. At least 2 applications are required
due to the absorbency of the wood.
10) When the resin is slightly cured (threads on the spatula), align the
ankle block and prosthetic foot with each other.
11) Insert the screw on the foot sole side in the prosthetic foot and
screw the ankle block to the prosthetic foot (mounting torque:
30 Nm).
12) Wipe off resin that comes out of the joint using a cloth.