NOTICE
Mechanical overload
Impaired functionality due to mechanical damage
Check the product for damage prior to each use.
Do not use the product if its functionality has been impaired.
Take any necessary measures (e.g. repair, replacement, inspection by the manufacturer's
customer service, etc.).
Signs of changes in or loss of functionality during use
Decreased forefoot resistance or changes in roll-over behaviour are noticeable indications of loss
of functionality.
4 Scope of delivery
Quantity
1
1
1
Spare parts/accessories (not included in the scope of delivery)
Fig.
1
Footshell
1
Insert for function module
5 Preparing the product for use
CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
Observe the alignment and assembly instructions.
INFORMATION
Not all of the materials may be available in your country. In this case, please contact the loc­
al branch of the manufacturer to obtain information on alternative materials.
5.1 Alignment
5.1.1 Applying/removing the footshell
Putting on the footshell
1) Slide the prosthetic foot into the footshell (see fig. 3, see fig. 4).
2) Press the heel of the prosthetic foot into the footshell until it engages (see fig. 5).
Removing the footshell
1) Push the lock of the footshell to the rear and pull up the heel of the prosthetic foot.
2) Remove the prosthetic foot from the footshell.
5.1.2 Bench Alignment
TT bench alignment
Required materials: 662M4 goniometer, 743S12 heel height measuring apparatus, 743A80 50:50 gauge,
alignment apparatus (e.g. 743L200 L.A.S.A.R. Assembly or 743A200 PROS.A. Assembly)
12
Designation

Instructions for use

Prosthetic foot
Protective sock
Reference number
Bench alignment process
Reference number
SL=Spectra-Sock-7
Reference number
2C1
2D11

Publicité