CAUTION
Turning the adjustment regulator too far
Injury due to faulty control or malfunction of the product.
► Do not turn the adjustment regulator further than the noticeable stop.
CAUTION
Slipping electrode
Injury due to unexpected product behaviour as a result of uncontrolled
activation of the prosthesis component.
► Inform the patient that uncontrolled movement of the prosthesis com
ponent may occur if the electrode slips.
CAUTION
Penetration of liquids
Injury due to faulty control or malfunction of the product.
► Make sure that no liquids can penetrate into the product.
► Do not use the product for bathing prostheses.
CAUTION
Establishing a connection between the skin and metal components
of the prosthesis with the use of carbon fibre
Injury due to unexpected product behaviour as a result of interference con
ducted by carbon fibres.
► During fabrication, make sure that no connection is formed between
the skin and metal prosthesis components by carbon fibres.
NOTICE
Improper care of the housing
Damage to the casing through the use of acetone, white spirit or similar
solvents.
► Only clean the housing with a damp cloth and mild soap (e.g.
453H10=1 Ottobock DermaClean).
NOTICE
Thermal overloading due to external heat supply
Destruction of the product.
► Do not expose the electrode to hot air or other sources of heat.
15