Important Information On The Inserts - Mectron Piezosurgery WHITE Mode D'emploi Et Maintenance

Masquer les pouces Voir aussi pour Piezosurgery WHITE:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

05.5

IMPORTANT INFORMATION ON THE INSERTS

WARNING:
- When the titanium nitride coating is
visibly worn out, the insert must be
replaced. Use of an overly worn out
EN
- Diamond-coated inserts: the diamond
inserts must be replaced when the
titanium nitride coating is visibly worn
out and, at any rate, after a maximum of
10 treatments.
- Do not activate the handpiece while
the insert is in contact with the treated
part, so that the electronic circuit can
recognize the best point of resonance
of the insert and allow its optimal
performance.
- Check the condition of wear of the
insert and that it is intact before and
during every use. In the event that a
drop in performance becomes apparent,
proceed to replace it.
- Use original Mectron inserts only. Use
of non-original inserts, in addition to
voiding the warranty, damages the
threading of the PIEZOSURGERY
handpiece, with the risk of no longer
being able to screw the original inserts
correctly during subsequent use.
Moreover, the device settings are tested
and guaranteed to operate correctly
only when original Mectron inserts are
used.
22
- Do not change the shape of the insert
This could cause it to break.
- Do not use an insert that has suffered
any type of deformation.
- Do not attempt to sharpen the insert
used.
- Always check that the threaded parts
of the insert and of the handpiece are
perfectly clean – see the Cleaning and
Sterilization Manual.
- If excessive pressure is applied on the
insert, it can cause the insert to break
and possibly harm the patient.
- For information on how to correctly use
the inserts, consult the annexed sheet
"Appropriate settings for the inserts
on the PIEZOSURGERY
you've purchased.
- Before using the PIEZOSURGERY
white, make sure you have prepared the
®
away the soft tissues, to avoid damaging
them. It may happen that, while cutting
the bone, accidental contact of certain
parts of the insert with the soft tissues
white" or the
®
®

Publicité

Table des Matières
loading

Table des Matières