Chapter 2 – Working with the sterile container
Inserting the indification plate and lead seal
After loading the sterile container, include an identification plate if necessary.
Slide the identification plate into the plate holder.
Fig. 3: Slide identification plate into the plate holder
After sealing the sterile container, insert a lead seal at the indicator lock of the closed container.
Fig. 4: Insert a lead seal
Sterilization
WARNING
Risk of contamination!
Sterilize with steam, taking note of the following: Sterilization with steam must be carried out with a
validated steam sterilization process (e.g. in a sterilizer in accordance with DIN EN 13060, EN 285,
ANSI/AAMI ST 46-2002 and validated in accordance with ANSI/AAMI/ISO 17665-1).
If sterilizing several sterile containers: Ensure that the maximum load capacity of the steam sterilizer,
as specified by the manufacturer, is not exceeded.
Ensure that the sterilization medium can reach all products stored in the sterile container.
10
Sterilize the sterile container (anodized surface) only with approved and validated sterilization
processes.
Sterilize expressly marked sterile container only with approved and validated sterilization
processes.