Chapter 2 – Working with the sterile container
Chapter 2 – Working with the sterile
container
First use
Thoroughly clean the new sterile container prior to first use, see
disinfection
After cleaning, use a suitable filter (see
Set-up
Removing the lid
The lid can be removed for cleaning the sterile container.
Open lid lock and remove the lid.
Changing the filter in the lid
Replace the filter at the following intervals, depending upon the filter type:
Disposable filter: prior to each sterilization cycle
PTFE permanent filter*: after max. 1 000 sterilizations
* For further information please consult the instructions for use of the permanent filter.
Changing the filter
1.
To open filter holder press back plastic latch.
2.
Remove filter holder once it is unlocked.
3.
Replace filter and install filter holder under latch noses.
4.
Push down filter holder so that it clicks into plastic latch.
Function checks
Visually inspect all components of the sterile container before each use to ensure there is no damage, and
check to ensure correct function:
Metal parts are not deformed.
Lid and bottom are not warped.
Filter holder sits in full surface contact.
Lid seals are fully intact (no cracks).
Plastic parts are fully intact (no cracks).
The permanent filter is undamaged.
The disposable filter has been changed.
The lock functions work properly (engages).
The filter holder latch functions work properly (engages).
The seal at filter holder is undamaged (no cracks).
Use sterile containers only if they are in mint condition. Replace any damaged components
immediately with original spare parts, or have the damaged components repaired.
8
[page 12].
Chapter 3 – Cleaning and
Changing the filter in the lid
[page 8]).