3-2 LCD display on the Control Unit Console
(A) Clock Display
(C) IMPLANT SYSTEM
(D) Memory Time
(F) TORQUE Graph
(H) Light Level
( I ) Coolant Level
Clock Display (Surgic Pro + ONLY)
(A)
Clock display. Clock is back upped by the internal battery.
(B)
SPEED
Selected speed indication. Bur rotation speed is displayed during rotation. When rotation stops, the maximum rotation
speed set is displayed. Depressing Foot Control: Rotation speed of the bur displayed on the LCD (Assuming the
correct gear ratio is selected) Foot Control speed varies depending on how far foot is depressed. Without pressing
Foot Control: Max Rotation preset setting displayed on the LCD
(C)
IMPLANT SYSTEM
The number of the implant system being used is displayed. (1 to 8)
Memory Time (Surgic Pro + ONLY)
(D)
Memory time remaining is displayed (Remain Time).
USB flash drive (Surgic Pro + ONLY)
(E)
When USB flash drive is inserted, it is displayed on the LCD.
(F)
TORQUE Graph
Display the actual operating torque range. When all bars illuminate, the operating torque is at preset torque. When
bars are half illuminated then the operating torque is approximately 50% of the preset torque.
(G)
TORQUE
Selected torque indication. When using the 1:1 direct drive or Speed increasing handpiece, the torque is not displayed.
(This function is appropriate for 20:1 reduction speed handpiece)
(H)
Light Level (For Optic ONLY)
Displays the selected light level; light levels are HIGH, LOW and OFF.
( I )
Coolant Level
Displays the selected coolant solution flow level; the selected Flow level is indicated by 0 to 5 levels of illuminated
indicators. No illumination indicates when the Coolant Solution Flow is off.
(J)
FWD/REV Direction
Displays the selected direction of handpiece rotation. When rotating, the circumference display rotates.
(K)
GEAR Ratio
Selected Gear Ratio of the handpiece.
(L)
PROGRAM Number
Selected Program Number indication. (1 to 8)
NOTICE
• Right after turning on the power source, the clock may display "88:88/00:00" or the LCD of the USB
REC key may blink, but these are not malfunctions.
• When the power source is turned on, the device will return to its default settings. To save a setting,
see 6-1 Programming the Micromotor Operation. (Default: Implant system 1, Light HIGH, Coolant Flow
3, Rotation F, Gear Ratio 20:1, Program Number 1, Torque 20N•cm, Speed 800min
• When packed for shipping, etc., the clock will blink on "00:00" if the clock has not been set. See 5-1 Clock and
Date.
• When the clock display blinks on "- -: - -," the clock batteries have run out. Contact the dealer where device was
purchased.
(E) USB flash drive
(B) SPEED
(J) FWD/REV Direction
(K) GEAR Ratio
(G) TORQUE
(L) PROGRAM Number
6